Anterior segment biometry during accommodation imaged with ultralong scan depth optical coherence tomography
2012-12-18 17:33点击:229次发表评论
作者:Du, C.ab, Shen, M.b, Li, M.c, Zhu, D.c, Wang, M.R.
机构: 浙江大学医学院附属第一医院眼科
期刊: OPHTHALMOLOGY2012年12月12期119卷

Purpose: To measure by ultralong scan depth optical coherence tomography (UL-OCT) dimensional changes in the anterior segment of human eyes during accommodation. Design: Evaluation of diagnostic test or technology. Participants: Forty-one right eyes of healthy subjects with a mean age of 34 years (range, 22-41 years) and a mean refraction of -2.5±2.6 diopters were imaged in 2 repeated measurements at minimal and maximal accommodations. Methods: A specially adapted and designed UL-OCT instrument was used to image from the front surface of the cornea to the back surface of the crystalline lens. Custom software corrected the optical distortion of the images and yielded the biometric measurements. The coefficient of repeatability and the intraclass correlation coefficient were calculated to evaluate the repeatability and reliability. Main Outcome Measures: Anterior segment parameters and associated repeatability and reliability upon accommodation. The dimensional results included central corneal thickness (CCT), anterior chamber depth and width (ACD, ACW), pupil diameter (PD), lens thickness (LT), anterior segment length (ASL = ACD + LT), lens central position (LCP = ACD + 1/2LT), and horizontal radii of the lens anterior and posterior surface curvatures (LAC, LPC). Results: Repeated measurements of each variable within each accommodative state did not differ significantly (P>0.05). The coefficients of repeatability (CORs) and intraclass correlation coefficients for CCT, ACW, ACD, LT, LCP, and ASL were excellent (1.2%- 3.59% and 0.998-0.877, respectively). They were higher for PD (18.90%-21.63% and 0.880-0.874, respectively) and moderate for LAC and LPC (34.86%-42.72% and 0.669-0.251, respectively) in the 2 accommodative states. Compared with minimal accommodation, PD, ACD, LAC, LPC, and LCP decreased and LT and ASL increased significantly at maximal accommodation (P<0.05), whereas CCT and ACW did not change (P>0.05). Conclusions: The UL-OCT measured changes in anterior segment dimensions during accommodation with good repeatability and reliability. During accommodation, the back surface of the lens became steeper as the lens moved forward. Financial Disclosure(s): The authors have no proprietary or commercial interest in any of the materials discussed in this article. © 2012 American Academy of Ophthalmology.

Department of Ophthalmology, First Affiliated Hospital, Zhejiang University, Hangzhou, Zhejiang, China

通讯作者:Wang, J.; Bascom Palmer Eye Institute, University of Miami, Miller School of Medicine, 1638 NW 10th Avenue, Miami, FL 33136, United States; email:jwang3@med.miami.edu
学科代码:眼科学   关键词:生物测量
来源: Scopus
Scopus介绍:Scopus 于2004年11月正式推出,是目前全球规模最大的文摘和引文数据库。Scopus涵盖了由5000多家出版商出版发行的科技、医学和社会科学方面的18,000多种期刊,其中同行评审期刊16,500多种。相对于其他单一的文摘索引数据库而言,Scopus的内容更加全面,学科更加广泛,特别是在获取欧洲及亚太地区的文献方面,用户可检索出更多的文献数量。通过Scopus,用户可以检索到1823年以来的近4000万条摘要和题录信息,以及1996年以来所引用的参考文献。数据每日更新。 马上访问Scopus网站http://www.scopus.com/
顶一下(0
您可能感兴趣的文章
    发表评论网友评论(0)
      发表评论
      登录后方可发表评论,点击此处登录